Korean Journal of Thoracic and Cardiovascular Surgery 2016; 49(5): 361-365  https://doi.org/10.5090/kjtcs.2016.49.5.361
Analysis of Sternal Fixation Results According to Plate Type in Sternal Fracture
Chun Sung Byun1, Il Hwan Park1, Wan Jin Hwang2, Yeiwon Lee2, and Hyun Min Cho3
1Department of Thoracic and Cardiovascular Surgery, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, 2Department of Thoracic and Cardiovascular Surgery, Konyang University Hospital, Konyang University College of Medicine, 3Department of Trauma Surgery, Pusan National University Hospital Trauma Center
Corresponding author: Hyun Min Cho, Department of Trauma Surgery, Pusan National University Hospital Trauma Center, 179 Gudeok-ro, Seo-gu, Busan 49241, Korea, (Tel) 82-51-240-7369 (Fax) 82-51-240-7719 (E-mail) csking1@daum.net
Received: September 1, 2015; Revised: November 2, 2015; Accepted: November 3, 2015.; Published online: October 5, 2016.
© The Korean Journal of Thoracic and Cardiovascular Surgery. All rights reserved.

cc This is an open access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properlycited.
Abstract

Background

Sternal fractures are relatively rare, and caused mainly by blunt anterior chest wall trauma. In most cases, sternal fractures are treated conservatively. However, if the patient exhibits problematic symptoms such as intractable chest wall pain or bony crepitus due to sternal instability, surgical correction is indicated. But no consensus exists regarding the most appropriate surgical method. We analyzed the results of surgical fixation in cases of sternal fracture in order to identify which surgical method led to the best outcomes.

Methods

We retrospectively reviewed the medical records of patients with sternal fractures from December 2008 to December 2011, and found 19 patients who underwent open reduction and internal fixation of the sternum with a longitudinal plate (L-group) or a T-shaped plate (T-group). We investigated patients’ characteristics, clinical details regarding each case of chest trauma, the presence of other associated injuries, the type of open reduction and fixation, whether a combined operation was performed, and postoperative complications.

Results

Of the 19 patients, 10 patients (52.6%) were male, and their average age was 56.8 years (range, 32 to 82 years). Seven patients (36.8%) had isolated sternal fractures, while 12 (63.2%) had other associated injuries. Seven patients (36.8%) were in the L-group and 12 patients (63.2%) were in the T-group. Three patients in the L-group (42.9%) showed a loosening of the fixation. In all patients in the T-group, the fracture exhibited stable alignment.

Conclusion

Open reduction and internal fixation with a T-shaped plate in sternal fractures is a safer and more efficient treatment method than treatment with a longitudinal plate, especially in patients with a severely displaced sternum or anterior flail chest, than a longitudinal plate.

Keywords: Sternum, Thoracic injuries, Fracture fixation, Bone plates
Introduction

Sternal fractures are relatively rare, accounting for <0.5% of total bone fractures and 8% to 10% of cases of blunt thoracic trauma [14]. Blunt anterior chest wall contusions, in which the chest collides with the steering wheel in traffic accidents, are the main cause of sternal fractures. Morbidity and mortality from sternal fractures are both directly related to the degree of co-injury, and have no clear relation to the sternal fracture itself [5,6]. Sternal fractures are treated conservatively in most cases. However, no reports have assessed the incidence of long-term complications after conservative treatment of sternal fractures. According to reports published over the last 50 years, if the internal organs sustain no damage in a severe sternal fracture, the mortality rate is low. No consensus exists regarding the most appropriate methods of surgical correction if a patient exhibits problematic symptoms such as intractable chest wall pain or bony crepitus due to sternal instability or a chest wall deformity [7,8]. Therefore, we analyzed the results of surgical fixation in patients with severe sternal fractures to determine the most appropriate surgical methods for sternal fractures.

Methods

We retrospectively reviewed the medical records of 117 patients with sternal fractures who were treated at Konyang University Hospital from December 2008 to December 2011. Of these patients, 19 patients underwent open reduction and internal fixation of the sternum. Patient characteristics such as age and gender, as well as information regarding the duration of admission, chest X-ray findings, chest computed tomography (CT) findings, and operative records, were collected. Clinical information including the mechanism of injury, time of injury, type of sternal fixation, co-injury site, surgical information for the co-injury site, and postoperative complications was also included. For patients with a sternal fracture, the surgical indications were complaints of severe pain and functional limitations with severe dislocation of the sternum in a chest X-ray or chest CT, or a flail motion that caused difficulty in weaning them from a mechanical ventilator.

Surgery was performed with a 10-cm mid-sternal incision between the sternal notch and the xyphoid process under general anesthesia. After dissection of the fractured sternum, the fractured site was corrected manually and fixed with a titanium plate (Hankil Tech Medical Co. Ltd, Hwaseong, Korea) with cortical screws (Hankil Tech Medical Co. Ltd). 7 patients were treated with a longitudinal plate that was occasionally reinforced by a stainless steelwire (the L-group) (Fig. 1A). 12 patients, a T-shaped plate without wiring was used (the T-group) (Fig. 1B). Following surgery, we checked for hemothorax or pneumothorax on an immediate postoperative chest X-ray. When patients became ambulatory, we confirmed sternal their fixation status via a lateral sternum X-ray. Incomplete fixation was defined as screw loosening in a radiologic examination after sternal fixation during the follow-up period.

Statistical analysis was performed using PASW SPSS statistics ver. 17.0.2 (SPSS Inc., Chicago, IL, USA). The Fisher exact test was used to evaluate differences between the groups, with p-values <0.05 considered statistically significant.

Results

Of the 19 patients who underwent surgical fixation of the sternum using a titanium plate and screws, 10 patients were male (52.6%), and their average age was 56.8 years (range, 32 to 82 years). The mean follow-up time was 9.2 months (range, 2 to 22 months), and no cases of mortality were associated with sternal fixation. Table 1 shows the general characteristics of the L-group (7 patients, 36.8%) and the T-group (12 patients, 63.2%). The average period from admission to surgery was 10.0 days for the L-group (range, 4 to 15 days) and 9.3 days for the T-group (range, 0 to 24 days). The average period from surgery to discharge was 26.8 days for the L-group (range, 9 to 53 days) and 15.3 days for the T-group (range, 7 to 34 days). A total of 7 isolated sternum fractures took place in 19 patients. The associated intrathoracic injuries included were large numbers of rib fractures, hemothoraces, hemopneumothoraces, pneumothoraces, and retrosternal hematomas. The associated extrathoracic injuries were clavicle fractures, facial bone fractures, liver lacerations, scapular fractures, and spine fractures. Open reduction and internal fixation of the ribs was concomitantly performed in 1 patient (14.3%) in the L-group and in 9 patients (66.7%) in the T-group. Operations to treat scapular fractures were performed in 2 patients (16.7%) in the T-group, and an operation to treat a clavicle fracture was performed in 1 patient (8.3%), also in the T-group. Postoperatively, pneumonia occurred in 1 patient in both groups, and acute respiratory distress syndrome (ARDS) occurred in 1 patient in the T-group. No postoperative mortalities occurred in either group. The patient who developed ARDS exhibited anterior flail chest, received mechanical ventilator care, and then progressed to ARDS; a tracheostomy was therefore performed 8 days after the sternal fixation.

When we compared the two groups defined according to plate type, differences were found in the frequency of isolated sternal fractures; 5 patients in the L-group (71.4%) did not have associated injuries, in comparison to 2 patients (16.7%) in the T-group. No significant differences were found in the fracture contour of the sternal fixation sites, the locations of sternal fixation, and the incidence of postoperative complications. However, there was a difference in the frequency of complete fixation between the two groups; 3 patients in the L-group (42.9%) underwent complete fixation, while all the T-group patients did so (Table 2). During the follow-up period, 2 patients had conservative treatment for incomplete fixation. One patient had the plate removed 8 months after surgery, and the sternum was fused completely after removal. In this patient, the anterior chest wall pain was much reduced and he experienced few limitations in terms of posture or daily activity.

Discussion

Surgical fixation of the sternum is performed in a very limited number of patients, because only few patients present with surgical indications of sternal fracture, such as severe dislocation of the fractured site or anterior flail motion. In our study, severe sternal fracture patients had numerous associated intrathoracic injuries, such as pneumothorax, hemothorax, and multiple rib fractures. Also, all patients’ cardiac enzyme levels were tested and, if elevated levels of any enzymes were found jointly with an electrocardiography abnormality, we performed echocardiography. All echocardiograms showed insignificant cardiac trauma in this study. Furthermore, all retrosternal hematomas were spontaneously absorbed after surgical fixation of the sternum. This implies that pulmonary complications such as pleural effusion, atelectasis, pneumonia, and ARDS occurred more frequently than cardiovascular injury in sternal fracture patients.

When early surgical fixation fails in patients with severe dislocations, a sternal fracture can progress to pseudoarthrosis, impaired ventilation, and transfiguration of the anterior chest wall. Early detection and treatment of sternal fractures is important for good quality of life after anterior chest wall trauma. For example, pseudoarthrosis after a sternal fracture can aggravate severe anterior chest wall pain leading to limitations in daily activity; additionally, delayed surgical correction of pseudarthrosis is very difficult for physician.

A non-invasive technique of sternal reduction is hyperextension of the thoracic spine. However, surgical experience in the treatment of severe sternal fractures has accumulated, and a variety of surgical techniques have been developed for sternal fractures. Prior to 2008, we treated most sternal fracture patients conservatively, even in cases of severe displaced fracture. However, since 2008, most displaced fractures have been treated with open reduction and internal fixation with titanium plate and screws. Many internal fixation methods exist for the treatment of sternal fractures, including a Steinmann pin for segmental fractures, stainless steel wiring for displaced fractures, and the use of a titanium plate and screws [914]. Especially in patients with severe dislocation of the sternal fracture site, a titanium plate and screws are more stable than figure-8 wiring. Titanium plates and screws for sternal fixation are both used for traumatic sternal fractures as well as chronic nonunion of the sternum after median sternotomy. Sternal fixation with a titanium plate and screws can more securely facilitate rapid bone union, and minimizes the dissection of the mediastinal soft tissue without causing damage to collateral blood vessels and the internal organs around the sternum. Furthermore, open reduction and internal fixation of a displaced sternal fracture can reduce anterior chest wall pain, improve pulmonary function, and prevent pseudoarthrosis and chest wall deformity.

In our study, the T-group patients exhibited more associated injuries and more oblique and segmental sternal fracture lines than the patients in the L-group; however, postoperative complications did not differ significantly between the two groups. This is because postoperative complications depend on the severity of the associated lung injury rather than the material and methods used in surgical fixation. In terms of fixation stability, the T-shaped plate was better than the longitudinal plate for internal fixation of the fractured sternum due to its low frequency of incomplete fixation. The use of a different plate type with similar surgical methods resulted in more incomplete fixations in the L-group, suggesting that T-shaped plates are better able to resist the tension-compression force in the sternum than longitudinal plates. However, the use of two longitudinal plates leads to more secure sternum fixation than the use of a single longitudinal plate [15]. Transverse tension resistance between each longitudinal plate is absent; in contrast, the T-shaped plate is more resistant against the tension-compression force from the sternum while maintaining the bone union of the fracture site [16]. The convex shape on the anterior side of the sternum is exposed to a tension force during deep inspiration, and the concave shape on the posterior side of the sternum increases the compression pressure. However, a T-shaped plate located on the anterior side of the sternum can overcome the bending force better than a longitudinal plate, and it reduces the compression pressure of the posterior side of the sternum to maintain contact with the fractured bone. In the early period of the sternal fixation procedures analyzed in this study, we only employed longitudinal plates; however, the T-shaped plate became preferred in the later period of this study. This change was attributable to the initial learning curve in open reduction and internal fixation on the part of the attending surgeons.

The present study was limited by its non-randomized and small sample size. In addition, although no significant differences were found in postoperative complication rates, the underlying general characteristics, trauma vectors, and associated injuries differed between the two groups, reducing the statistical power of the study. However, this study showed the results of sternum fracture fixation according to plate type, which can help guide sternum fracture management strategies in the future.

In conclusion, open reduction and internal fixation with a T-shaped plate in sternal fractures is a safer and more efficient method than fixation with a longitudinal plate, especially in patients with a severely displaced sternum or anterior flail chest. This conclusion needs to be verified by additional prospective studies with larger populations, in order to provide clinically relevant findings regarding the short-term results of bone union and the long-term functional and cost advantages of fixation with a T-shaped plate.

Acknowledgments

This study was supported by a Grant of the Samsung Vein Clinic Network (Daejeon, Anyang, Cheongju, Cheonan; Fund No. KTCS04-057).

Figures
Fig. 1. Open reduction and internal fixation of sternum with (A) longitudinal plate with stainless steel wiring and (B) T-shaped plate.
Tables

General characteristics of patients (n=19)

Characteristic Longitudinal plate group (n=7)  T-shaped plate group (n=12) 
Sex (male)3 (42.9)7 (58.3)
Mean age (yr)60.9±13.9 (41–79)54.5±15.3 (32–82)
Mean follow-up (mo)9.8±6.3 (8–22)8.6±4.8 (2–17)
Duration from admission to surgery (day)10.0±3.8 (4–15)9.3±7.3 (0–24)
Duration from surgery to discharge (day) 26.8±14.2 (9–53)15.3±10.9 (7–34)
Injury mechanism
 In-car TA6 (85.7)6 (50.0)
 Pedestrian TA04 (33.3)
 Fall01 (8.3)
 Assault1 (14.3)1 (8.3)
Thoracic injuriesa)
 Rib fracture2 (28.6)7 (58.3)
 Traumatic hemothorax2 (28.6)3 (25.0)
 Traumatic hemopneumothorax1 (14.3)2 (16.7)
 Traumatic pneumothorax1 (14.3)1 (8.3)
 Retrosternal hematoma1 (14.3)4 (33.3)
Associated injuriesa)
 Liver laceration1 (14.3)1 (8.3)
 Scapular fracture02 (16.7)
 Facial bone fracture1 (14.3)2 (16.7)
 Clavicle fracture01 (8.3)
 Spine fracture02 (16.7)
Combined operationsa)
OR and IF of ribs1 (14.3)8 (66.7)
OR and IF of scapula02 (16.7)
OR and IF of clavicle01 (8.3)
Postoperative complications
 Pneumonia1 (14.3)1 (8.3)
 Acute respiratory distress syndrome01 (8.3)
Late mortality00

Values are presented as numbers (%) or mean±standard deviation (range).

TA, traffic accident; OR, open reduction; IF, internal fixation.

a)Multiple responses.

Comparison of surgical parameters between the longitudinal plate and T-shaped plate groups (n=19)

Characteristic Longitudinal plate group (n=7)  T-shaped plate group (n=12)  p-value 
Type of sternal fracture0.045
 Isolated5 (71.4)2 (16.7)
 Combined injuries2 (28.6)10 (83.3)
Fracture contour of sternal fixation site 0.364
 Linear7 (100.0)8 (66.7)
 Oblique02 (16.7)
 Segmental02 (16.7)
Location of sternal fixation0.850
 Manubrium01 (8.3)
 Sternal angle02 (16.7)
 Upper body3 (42.9)2 (16.7)
 Mid-body3 (42.9)5 (41.7)
 Lower body1 (14.3)2 (16.7)
Postoperative complications0.523
 Yes1 (14.3)2 (16.7)
Result of surgical fixation0.036
 Complete4 (57.1)12 (100.0)
 Incomplete3 (42.9)0

Values are presented as numbers (%). Group comparisons were analyzed by the Fisher exact test.

References
  1. Potaris, K, Gakidis, J, Mihos, P, Voutsinas, V, Deligeorgis, A, and Petsinis, V (2002). Management of sternal fractures: 239 cases. Asian Cardiovasc Thorac Ann. 10, 145-9.
    Pubmed CrossRef
  2. Mayba, II (1985). Non-union of fractures of the sternum. J Bone Joint Surg Am. 67, 1091-3.
    Pubmed
  3. Helal, B (1964). Fracture of the manubrium sterni. J Bone Joint Surg Br. 46, 602-7.
    Pubmed
  4. Brookes, JG, Dunn, RJ, and Rogers, IR (1993). Sternal fractures: a retrospective analysis of 272 cases. J Trauma. 35, 46-54.
    Pubmed CrossRef
  5. Gouldman, JW, and Miller, RS (1997). Sternal fracture: a benign entity?. Am Surg. 63, 17-9.
    Pubmed
  6. Trinca, GW, and Dooley, BJ (1975). The effects of mandatory seat belt wearing on the mortality and pattern of injury of car occupants involved in motor vehicle crashes in Victoria. Med J Aust. 1, 675-8.
    Pubmed
  7. Sadaba, JR, Oswal, D, and Munsch, CM (2000). Management of isolated sternal fractures: determining the risk of blunt cardiac injury. Ann R Coll Surg Engl. 82, 162-6.
    Pubmed KoreaMed
  8. Harley, DP, and Mena, I (1986). Cardiac and vascular sequelae of sternal fractures. J Trauma. 26, 553-5.
    Pubmed CrossRef
  9. Van Sterkenburg, SM, Brutel de la Riviere, A, and Vermeulen, FE (1990). Sternal fixation with resorbable suture material. Eur J Cardiothorac Surg. 4, 345.
    Pubmed CrossRef
  10. Kitchens, J, and Richardson, JD (1993). Open fixation of sternal fracture. Surg Gynecol Obstet. 177, 423-4.
    Pubmed
  11. Molina, JE (2005). Evaluation and operative technique to repair isolated sternal fractures. J Thorac Cardiovasc Surg. 130, 445-8.
    Pubmed CrossRef
  12. Hills, MW, Delprado, AM, and Deane, SA (1993). Sternal fractures: associated injuries and management. J Trauma. 35, 55-60.
    Pubmed CrossRef
  13. Bonney, S, Lenczner, E, and Harvey, EJ (2004). Sternal fractures: anterior plating rationale. J Trauma. 57, 1344-6.
    Pubmed CrossRef
  14. Athanassiadi, K, Gerazounis, M, Moustardas, M, and Metaxas, E (2002). Sternal fractures: retrospective analysis of 100 cases. World J Surg. 26, 1243-6.
    Pubmed CrossRef
  15. Schulz-Drost, S, Oppel, P, and Grupp, S (2015). Surgical fixation of sternal fractures: preoperative planning and a safe surgical technique using locked titanium plates and depth limited drilling. J Vis Exp, e52124.
    Pubmed KoreaMed
  16. Al-Qudah, A (2006). Operative treatment of sternal fractures. Asian Cardiovasc Thorac Ann. 14, 399-401.
    Pubmed CrossRef